Page 26 Leica EM CPD030 Operating Manual 09/09
5.11 Removing dried specimen
Close the shut-off valve of the C02 bottle.
Open the screw-on cover of specimen pressure chamber.
Remove specimen holder with dried specimen from pressure chamber for further
processing.
Since the dried specimen is highly hygroscopic, it has to be coated as quickly as
possible with a thin metal- or carbon film to protect it from atmospheric humidity.
If this is not possible, it is recommended to keep the specimen in a desiccator until it
is processed further.
5.12 Switching off the unit
Leave the specimen chamber open. Leave the C02 bottle closed.
Press the MEDIUM IN key. The gas contained in the connection tube escapes
through the open specimen pressure chamber.
Press the MEDIUM IN key again. This will close the medium-inlet valve.
Close the specimen chamber with screw-on cover.
Switch off the main switch located on the unit.